PURPOSE: To evaluate methods of measuring anterior chamber depth (ACD) before and after cataract surgery. SETTING: St. Erik's Eye Hospital, Stockholm, Sweden. METHODS: A-scan (BVI Axis) and Scheimpflug imaging (Nidek EAS-1000) were used to measure ACD preoperatively and 6 and 18 weeks after phacoemulsification with implantation of a poly(methyl methacrylate) (PMMA) intraocular lens (IOL) in 23 patients. Because of a large systematic difference between the 2 methods, measurement with Orbscan (Orbtek) and optical pachymetry (Haag-Streit) were included when measurements were repeated 36 weeks postoperatively. A t test for paired observations was used for statistical analysis. RESULTS: Preoperatively, the mean A-scan measurements were significantly shorter than the Scheimpflug values: 3.05 mm +/- 0.36 (SD) and 3.37 +/- 0.35 mm, respectively (P