Results of 884 first-time shunts inserted in the time period from 1958 to 1989 are retrospectively evaluated, 1) to perform a durability analysis of a shunt based on Kaplan-Meyer method, 2) to compare the rate of revision for ventriculo-atrial (VA) and ventriculo-peritoneal (VP) shunts, 3) to compare the durability of a VA shunt with a VP shunt and 4) to do a stratified durability analysis comparing the VA and VP shunts in relation to the following background variables: shunt type, time period and age of the patient. Furthermore the specific complications related to VA and VP shunts are identified based on findings in the literature. Overall one-year shunt durability is 57% and five-year shunt durability is 37%. The median shunt durability is 1.68 years. Revision rate is 51% for VA shunts and 38.5% for VP (p