Compared with the first-generation Arctic Front cryoballoon (ARC-CB), the new Arctic Front Advance cryoballoon (ARC-Adv-CB) increases the efficient CB-tissue contact surface during freezing, which may increase the incidence of phrenic nerve (PN) palsy (PNP).
To evaluate the safety and efficacy of paroxysmal atrial fibrillation (AF) ablation with the ARC-Adv-CB as well as the merits of a predictor of PNP.
AF ablation was performed by using a "single 28-mm big CB" approach. The rate of pulmonary vein (PV) isolation with a first cryoapplication was measured. The distance between the CB and a PN pacing catheter in the superior vena cava was measured to predict PNP during freezing.
In 147 patients, PV were isolated with a single cryoapplication in 205 (81.3%) of 252 PV treated with the ARC-CB and in 280 (90.3%) of 310 PV treated with the ARC-Adv-CB (P = .003). The mean time to PV isolation was 52 ± 34 seconds and 40 ± 25 seconds (P