National surveys of more than 600 facilities that performed chest, lumbosacral spine, and abdominal examinations were conducted as a part of the Nationwide Evaluation of X-Ray Trends program. Radiation exposures were measured with use of a set of standard phantoms developed by the Center for Devices and Radiological Health of the Food and Drug Administration, U.S. Public Health Service. X-ray equipment parameters, film processing data, and data regarding techniques used were collected. There were no differences in overall posteroanterior chest exposures between hospitals and private practices. Seventy-six percent of hospitals used grids, compared with 33% of private practices. In general, hospitals favored a high tube voltage technique, and private facilities favored a low tube voltage technique. Forty-one percent of private practices and 17% of hospitals underprocessed their film. Underprocessing in hospitals increased from 17% in 1984 to 33% in 1987. Average exposure values for these examinations may be useful as guidelines in meeting some of the new requirements of the Joint Commission on Accreditation of Healthcare Organizations.